xMAP Technology Create Template
PN 89-00002-00-084 Rev. B 21
Qualitative Assay
Template
You define each Qualitative template using a single product. The
product must contain one standard. The standard is assigned a
Drain to reservoir Removes and refills sheath fluid from the cuvette.
Sheath fluid is expelled into the reservoir.
Drain to well Removes and refills sheath fluid from the cuvette.
Sheath fluid is expelled into a well.
Prime Primes the system to remove air from the sheath
side of the analyzer.
Sanitize Acquires sanitizing solution from the reservoir to
sanitize or eliminate air in the cuvette.
Skip Program a well skip into template.
Soak from
Reservoir
Soaking the probe replaces sheath fluid in the
probe with distilled water. You should perform the
Soak command at the end of each day. Place at
least 2 mL of distilled water in the reservoir.
Soak from Well Soaking the probe replaces sheath fluid in the
probe with distilled water. You should perform the
Soak command at the end of each day. Place at
least 275 µL of distilled water in the well.
Verify CON 1 Acquire CON 1 beads from a well to verify that the
system is working within the calibrated settings.
Verify CON 2 Acquire CON 2 beads from a well to verify that the
system is working within the calibrated settings.
Warmup Checks for proper system warm up and if needed
begins the warmup process.
Wash from
Reservoir
Acquires wash solution from reservoir to wash the
system. Place at least 2 mL of solution in the
reservoir.
Wash from Well Acquires wash solution from a well to wash the
system. Place at least 200 µL in the well.
XY Heater Off Turns the heater off.
XY Heater On Sets the indicated temperature (XyHtrTemp), turns
the heater on, and continues to the next command.
XY Heater On
(wait)
Sets the indicated temperature (XyHtrTemp), turns
the heater on, waits for the heater to stabilize at
the heater set point (with a 30 minute timeout), and
then continues to the next command.
Table 3: Available Commands (Continued)
Command Definition