Luminex 200 System User Manual Manual de usuario

Busca en linea o descarga Manual de usuario para Equipo Luminex 200 System User Manual. Luminex 200 System User Manual User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 63
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
Luminex
®
200
System
User Manual
For In Vitro Diagnostic Use
Vista de pagina 0
1 2 3 4 5 6 ... 62 63

Indice de contenidos

Pagina 1 - User Manual

Luminex® 200™ SystemUser ManualFor In Vitro Diagnostic Use

Pagina 2

5. TERM AND TERMINATION. Your rights under this EULA are effective until termination.You may terminate this EULA at any time by destroying the SOFTWAR

Pagina 3

Table of ContentsChapter 1 Safety ...1I

Pagina 4 - 2. Warranties

Chapter 3 Maintenance and Cleaning ...21Daily Maintenance ...

Pagina 5

Please become familiar with the information in this chapter before using the equipment. Donot perform procedures on your Luminex® 200™ system that are

Pagina 6

SymbolsThese symbols describe warnings, cautions, and general information used in the operation ofthis instrument. Some of these symbols are further d

Pagina 7 - 5. Patent Disclaimer

impaired or the warranty voided if the system is used in a manner not specified by theinstructions or by Luminex Corporation.FluidicsThis system conta

Pagina 8 - Software

FIGURE 1.Laser Radiation Caution LabelWARNING: Under NO circumstances should you remove the Luminex 200analyzer cover! When performing routine mainten

Pagina 9

FIGURE 2.Laser Class LabelWARNING: Use of controls or adjustments or performance of proceduresother than those specified herein may result in hazardou

Pagina 10 - For In Vitro Diagnostic Use

MechanicalWARNING: During operation, this system contains exposed, moving parts.Risk of personal injury is present. Observe all warnings andcautions.W

Pagina 11 - Table of Contents

Decontaminating the Luminex 200 Analyzer for ReturnShipmentLuminex Technical Support will give you a Return Material Authorization (RMA) number ifthey

Pagina 13 - Chapter 1: Safety

Luminex® 200™ System User Manual8For In Vitro Diagnostic Use

Pagina 14 - Safety Precautions

Theory of OperationLuminex xMAP technology is based on flow cell fluorometry with Luminex-developedinnovations. The fluidics, optics, robotics, temper

Pagina 15 - Luminex 200 Analyzer Laser

through the Luminex 200 analyzer. Results of the analyses are processed and provided in areport format.HardwareThe Luminex 200 system includes the fol

Pagina 16 - Laser Radiation Caution Label

Required Laboratory Reagents• Household bleach• 70% isopropanol or 70% ethanol• Mild detergent• Distilled waterLuminex SoftwareLuminex software provid

Pagina 17 - Barcode Reader Laser

• Internal sample carry over: < 0.9%• Soluble background fluorescence emission at 575 nm automatically subtracted fromfluorescence intensity values

Pagina 18 - Blue Indicator Light

• Shipping and storage: The allowable shipping and storage temperature and humidityranges are 0°C to + 50°C and 20-80% noncondensing, respectively• In

Pagina 19 - Shipment

Luminex SD System General• Ambient temperature: 15°C to 30°C (59° to 86°F)• Humidity: 20% to 80%, noncondensing• Altitude: designed to operate at up t

Pagina 20 - System User Manual

VortexUse VWR product number 58816-12, with a speed range of 0 to 3200 rpm, or equivalent.Bath SonicatorUse Cole-Parmer® product number 08849-00, with

Pagina 21 - Chapter 2: The System

FIGURE 3.Back of the Luminex 200 Analyzer and Luminex XYPInstrument1 Air intake filter access door 5 Communication Ports (DB9)2 Power Switch 6 XYP Com

Pagina 22 - Hardware

FIGURE 4.Fluidics Components1 Cheminert Fitting 3 Front thumbscrew2 Probe Holder 4 Luminex XYP Instrument Sample ProbeAccess DoorsThe Luminex 200 anal

Pagina 23 - Luminex Software

© Luminex Corporation, 2011. All rights reserved. No part of this publication may bereproduced, transmitted, transcribed, or translated into any langu

Pagina 24 - Luminex 200 Analyzer General

FIGURE 5.Luminex 200 Analyzer Access Doors1 Left door, access to service panel 2 Center door, access to syringeAir Intake FilterA replaceable air inta

Pagina 25 - Electronics

WARNING: The waste container should not be placed on top of theinstrument. Ensure that the waste tubing is not elevated abovethe level of the Luminex

Pagina 26 - PC Specifications

Luminex® 200™ System User Manual20For In Vitro Diagnostic Use

Pagina 27 - System Overview

To ensure accurate test results, properly clean and maintain the Luminex 200 system. Readand follow all instructions in this section. For your conveni

Pagina 28

Adjust the Sample Probe Vertical HeightAdjust the sample probe vertical height each time you change the type or style of microtiterplate.1. Remove the

Pagina 29 - Access Doors

Sheath and Waste FluidsReplace the sheath fluid and empty the waste container as required. Be careful not to touchthe seal in the waste bottle cap or

Pagina 30

Clean the Sample ProbeWARNING: Be sure that the system is not performing an operation whenyou remove the sample probe.CAUTION: The Luminex analyzer sa

Pagina 31

Clean the Sample ProbeWARNING: Be sure that the system is not performing an operation whenyou remove the sample probe.CAUTION: The Luminex analyzer sa

Pagina 32

3. Grasp the tubing and pull the filter 3 to 4 inches from the unit. See the following figure. FIGURE 6.Grasping the Tubing4. Remove the filter with o

Pagina 33 - Daily Maintenance

Luminex XYP Instrument Air Intake Filter1. Disconnect the Luminex XYP instrument from AC power by turning off the power switchon the rear of the Lumin

Pagina 34 - Routine Tasks

Standard Terms and Conditions for Use of InstrumentProductBy opening the packaging containing this product ("Product") or by using such Prod

Pagina 35 - Visual Inspection

Syringe SealWARNING: The syringe arm does not deactivate when changing theplunger; injury could result if the system is not unplugged.1. Turn the powe

Pagina 36 - Clean Exterior Surfaces

Replace the SyringeTo replace the syringe:1. Screw the syringe into place.2. Grab the base of the syringe plunger and gently pull down on it until it

Pagina 37 - Every Six Months

AnnuallySheath Filter1. Disconnect the Luminex 200 analyzer from AC power by turning off the power switch onthe rear of the analyzer, then unplugging

Pagina 38 - Grasping the Tubing

1. Turn off the power switch on the rear of the analyzer or instrument, then unplug theinstrument power cord from the wall source. Remove the power co

Pagina 39 - Removing the Screen

9. If the pressure is equal to the value noted in step 2 (within +/-0.1psi), then the set up iscomplete. If not, continue with the next step. After ea

Pagina 40 - Syringe Seal

Luminex 200 Maintenance LogsMonth:Year:Use this form to record information over a four-week period. Fill in the month(s) and yearabove. Fill in the da

Pagina 41 - Replace the Syringe

Loosensheath capTurn offsystem(optional)TABLE 2.Longer Term MaintenanceWEEKLYVisual inspection Date/Initials: Date/Initials: Date/Initials: Date/Initi

Pagina 42 - As required

Troubleshooting procedures help users isolate, identify, and remedy problems with theLuminex 200 analyzer and Luminex XYP. This chapter does not troub

Pagina 43 - Opening the Module Door

Symptom Possible problem SolutionAnalyzer will not turn on, or XYPwill not turn on.The power cord is disconnected. Verify that the power cord isplugge

Pagina 44 - Storing the System

Symptom Possible problem SolutionThere is a leak in the system. Check for system leaks. A leakwill be obvious if there is fluid onthe surface on which

Pagina 45 - Luminex 200 Maintenance Logs

hardware not provided by Luminex. If Product is purchased from a Luminex authorizedreseller, any warranty obligations shall be provided in writing dir

Pagina 46 - Longer Term Maintenance

Symptom Possible problem SolutionFluid dripping from the sampleprobe.The sample probe is clogged Clean the sample probe. Seeinstructions.The sample th

Pagina 47 - 200 System

Symptom Possible problem SolutionSample arm is stuck in the downposition.The sample probe height is toolow, or the path to the well isblocked.DO NOT t

Pagina 48 - Pressurization

Symptom Possible problem SolutionCalibration is slow or fails. The calibration microspheresare not fully suspended.Vortex the calibration vials toresu

Pagina 49 - Fluid Leaks

Symptom Possible problem SolutionLaser-related issue. Verify fluid is moving through thesystem by performing a Wash,which causes fluid to go out towas

Pagina 50 - Sample Probe

Symptom Possible problem SolutionThe waste line was movedduring system operation,resulting in unstable flow rate.Verify that the waste line is notmove

Pagina 51

Symptom Possible problem SolutionThe acquisition volume is set toohigh.Set the acquisition volume to atleast 25 µL less than the actualvolume in your

Pagina 52

FIGURE 13.Normal Bead DetailThe histogram for the bead detail above looks like this:FIGURE 14.Normal HistogramLuminex® 200™ System User Manual44For In

Pagina 53

Symptom Possible problem SolutionxMAP microspheres classify toohigh.You may be usingphotobleached calibrationmicrospheres.Replace the calibrationmicro

Pagina 54 - Acquisition Problems

Symptom Possible problem SolutionThe solvent is incompatible. View a list of incompatiblesolvents at the LuminexTechnical Support website -http://lumi

Pagina 55 - Bead Detail Irregularities

6. Press the Prime button on the front panel of the SD system.7. Sheath fluid will pump from the Sheath Delivery System reservoir into the 20L SheathB

Pagina 56 - Normal Histogram

indicated on the Product label, the Product has not received approval from the United StatesFood and Drug Administration or other federal, state or lo

Pagina 57

Luminex® 200™ System User Manual48For In Vitro Diagnostic Use

Pagina 58 - Luminex SD Problems

HardwareNOTE: These part numbers are subject to change without notice.NOTE: Common description of product is enclosed in brackets.Product Description

Pagina 59

Product Description Customer NumberPower Cord, USAPower Cord, AustraliaPower Cord, BrazilPower Cord, DenmarkPower Cord, Germany, Sweden, France,Belgiu

Pagina 60

Product Description Customer NumberMicrospheres, LX100, CL1 CL2 Calibration (xMAP ClassificationCalibrator)L100-CAL1Microspheres, LX100, RP1 Calibrati

Pagina 61 - Chapter 5: Product Numbers

contract or any other theory of law or equity arising out of, directly or indirectly, the use of theProduct or by reason of Buyer's failure to pe

Pagina 62 - Software

End-User License Agreement (EULA) for Luminex®xPONENT® SoftwareThis Luminex End-User License Agreement (“EULA”) is a legal agreement between you(eithe

Pagina 63 - Product Numbers

3. GRANT OF LICENSE. Subject to the terms and conditions of this EULA, Luminex herebygrants to you a nonexclusive, nontransferable, nonassignable lice

Comentarios a estos manuales

Sin comentarios