Luminex xPONENT 4.2 for FLEXMAP Manual de usuario

Busca en linea o descarga Manual de usuario para Software Luminex xPONENT 4.2 for FLEXMAP. Luminex xPONENT 4.2 for FLEXMAP User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 148
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente

Indice de contenidos

Pagina 1 - FLEXMAP 3D

xPONENT® forFLEXMAP 3D®Software Version 4.2 User Guide

Pagina 2

xPONENT® for FLEXMAP 3D®x

Pagina 3 - 2. Warranties

12. Highlight a range and click Delete Range to delete the range.13. Click OK to apply the new settings to the analyte first clicked, or Apply to All

Pagina 4

• Replicate Count - Defines a quantity of replicate sets from one to nine.• Grouping - Selects the sequence in which the replicates are laid out in pl

Pagina 5

• Pre Batch Routine - Opens the Commands and Routines dialog box so you can select apre-batch command or routine.• Plate Name - Specifies the plate yo

Pagina 6 - 5. Patent Disclaimer

Use this tab to select or edit analytes used in the batch or protocol. This tab contains thefollowing:This tab contains the following:• Analytes grid

Pagina 7 - Software

sets that are not in the sample. Therefore, choose only the bead sets present in yoursample.• Apply All - Applies the information in the Units and Cou

Pagina 8

• Units - The unit of measurement you specified in the Unit box. Click this box to type avalue for the analyte.• Count - Type the desired bead count f

Pagina 9

After you select from this box, the Std/Ctrl Details tab opens so you can create a new lotor kit.• Installed Kits And Lots - Displays information abou

Pagina 10 - for FLEXMAP 3D

Use the Maintenance page to maintain and calibrate your system. This page contains thefollowing tabs:• Auto Maint• Lot Management• Cmds & Routines

Pagina 11 - Table of Contents

This tab contains routines to initialize, verify, and maintain the analyzer.Automated Maintenance Options - Provides buttons for commonly used mainten

Pagina 12

Running Calibration and VerificationBefore running Calibration/Verification from the Auto Maint tab, you need to import Cal andVer kit information. Pe

Pagina 13

Table of ContentsChapter 1 Introduction ...1Safet

Pagina 14

Active Reagents - Use this section to change calibration or performance verification kits,lot numbers and expiration dates, or to delete a kit.• Calib

Pagina 15 - Chapter 1: Introduction

• Verifier - If a verification reagent is selected, the Lot Details table displays targetinformation.• Sheath Fluid - Click Edit, enter the Current Lo

Pagina 16

• Remove Air Bubbles (Luminex)• Remove Clog (Luminex)• Revive After Storage (Luminex)• Verification (Luminex)• Weekly Maintenance (Luminex)Commands -

Pagina 17 - Screen Elements

NOTE: You should perform a Self Test as part of your scheduled weeklymaintenance.Click a command to add it to a new routine, to modify an existing rou

Pagina 18 - System Monitor

6. Click Save As to save the new routine. The Save Routine dialog box opens.7. Type the routine name in the Routine Name box, then click OK.Editing a

Pagina 19 - Quick Start

2. Click Import.3. In the Open dialog box, browse for the file you want to import, then click Open. Theroutine becomes the active routine.Exporting a

Pagina 20

• Plate Location pin - A green pin that shows the well location used to auto adjust the probeheight for the main plate, reservoir, and/or strip wells.

Pagina 21

8. In the Reservoir section, click in an off-plate position.9. In the Strip-Wells section, click on a well location.10. Click Auto Adjust Height. The

Pagina 22 - Basic Procedures

Use this tab to view information and diagnostics about the Luminex instrument.This tab contains the following information:• Software• Version• Operati

Pagina 23

• System Temperature• DD Temperature• CL1 Temperature• CL2 Temperature• CL3 Temperature• CL1 Voltage• CL2 Voltage• CL3 Voltage• DD Voltage• RP1 Voltag

Pagina 24 - Initial Startup

Importing a Batch ...49Ex

Pagina 25

• Log Date• Message• Code• Export - Opens the Export System Log dialog box. Select a name and location for thesystem log export and click OK. Select O

Pagina 26 - System Initialization

Support Utility TabMaintenance > Support UtilityUse this tab to type information that will be useful in the event you need to contact LuminexTechni

Pagina 27 - Logging Off and Exiting

• Clear - Clears all the information in the Name, Phone, Company, Email, and Commentboxes.• Browse - Opens the Browse for Folder dialog box to choose

Pagina 28 - Using Online Help

You must have the appropriate license to view some of the features of in the Admin page.This page contains the following tabs:• System Setup• Group Se

Pagina 29 - Software Packages

Use this tab to configure system settings such as application settings, LIS settings,maintenance page options, and external analysis program settings.

Pagina 30

External Analysis Program SettingsThis option is for those using a program other than xPONENT® to analyze collected data.• Installed Analysis Programs

Pagina 31

• System Initialization Procedure - Displays the available selections to use as the defaultsystem initialization procedure. After selecting an option,

Pagina 32 - Sample Volume

NOTE: The Home page, and in some instances the Admin page, cannot bemoved.To arrange the main navigation buttons:1. Select or clear the check boxes by

Pagina 33

2. Click System Setup. 3. Under Maintenance Options, you can select a procedure from the drop-down list. • Laser warm-up, fluidics prep, calibration,

Pagina 34

Group Profile - The following user groups are predefined:• Administrator• Supervisor• Service• Technician2• Technician1• ReviewerThe user will belong

Pagina 35 - Chapter 3: Samples Page

Commands and Routines ...97Crea

Pagina 36 - Creating New Samples

• Validate and Invalidate Results• Approve Batch• Reanalyze Results• Save Batch after changing results• Change Formula• Reacquire errored wells for pa

Pagina 37

You must have administrative privileges and be using the Secure version of xPONENT or the21 CFR Part 11 version to perform the following tasks.• Creat

Pagina 38 - Editing a Sample List

Restoring Account StatusIf users attempt to log in unsuccessfully more than the number of times allowed, they will belocked out. In the User list, cli

Pagina 39

• Technician2• Technician1• Reviewer• Password - Type a unique password for the user in this box.• Reenter Password - This becomes active when you typ

Pagina 40

• Group Profile - The user's group or category.• Change Password After Next Login - A box that, when checked, indicates that the userwill change

Pagina 41 - Chapter 4: Batches Page

• Allow running a batch if the instrument is not warmed up - Allows batches to run if theinstrument is not warmed up.• Allow running or saving a batch

Pagina 42 - Setting Up Batches

• Error (Stop) - Stops the batch.• Warning ( Pause) - Pauses the batch.• Warning (Log) - Logs a warning if the number of successive wells in the Well

Pagina 43 - Using the Batches Page

• Scheduled Maintenance - A scheduled maintenance routine is now due.• Low Bead Count Detected - A bead count below the accepted range is encountered.

Pagina 44 - Protocol Subtab

4. If you choose to send an email notification for a specific event, and the email address isnot already set up, click Setup Email to open the Setup E

Pagina 45

CSV and Batch Export OptionsThis tab contains the following:• Automatically export results CSV file when batch is complete - Automatically exportthe r

Pagina 46

xPONENT® for FLEXMAP 3D®xiv

Pagina 47

• Automatically convert the raw run files to CSV format for each well in the batch -Click the check box if you want the automatic conversion.• Cancel

Pagina 48

The Archive Utility can be used within xPONENT to set up frequency and type of files toback up. Statistics about disk space and archive space used, as

Pagina 49

• System Restore - Click this to restore your system. It requires a log in and then refers youto the Start menu to restore your system.• Manual Archiv

Pagina 50

Setting Up Automatic ArchivingClick Edit in the Schedule Overview box to enable the fields to accept your edits and toenable the Browse button. The la

Pagina 51 - Settings Subtab

1. In the Archive Utility, click Manual Archive to open the Manual Archive window. 2. Using the tabs on the left side of the window, select the type o

Pagina 52

Contact Luminex Technical Support to upgrade xPONENT® for FLEXMAP 3D® or to obtain anew license.NOTE: You must restart the computer for the new licens

Pagina 53 - Analytes Tab

Use the Schedule tab to enable or change the recurrence of scheduled maintenancereminders.• Select the time of day that you want to receive alerts in

Pagina 54

2. On the Schedule tab, use the drop-down menus to edit the following items for anyscheduled activity: • Alert Time - the time of day that you want to

Pagina 55

1. Type a company name in the Company box, and any additional information in the Info:box. NOTE: The logo file should be 920 x 125 pixels. If you want

Pagina 56

Safety PrecautionsDANGER: Samples and waste fluid can contain biohazardous infectiousagents. Handle them at Biosafety level 2, as recommended forany p

Pagina 57

The Home page displays a Welcome message, batch creation buttons, Daily Activitiesshortcuts, and the Installed Protocols list.Return to the Home page

Pagina 58

• System Initialization - Opens the routine specified in the Maintenance Options on theAdmin page.• Shutdown - Opens the System Shutdown command in th

Pagina 59

Right-Click MenuCertain sections of the software such as tables, lists, and text boxes have right-click optionmenus. Menus are different depending upo

Pagina 60 - Create a New Multi-Batch

Command Display - Displays the following:• The command currently running.• The system state (i.e. running, idle, etc.).• Date and time.Progress - Disp

Pagina 61 - Creating a New Multi-Batch

© Luminex Corporation, 2012. All rights reserved. No part of this publication may bereproduced, transmitted, transcribed, or translated into any langu

Pagina 62

To Go to Expanded HelpAdjust the sample probe height Home > Probe and Heater Adjusting the Sample ProbeInitialize the system Home > System Initi

Pagina 63 - Batch Procedures

• Serial Number• Firmware Version• XYP Heater Temperature• Sheath Pressure• Calibration/Verification Status• Last F3DCAL1 Calibration• Last F3DeCAL1 C

Pagina 64 - Deleting a Batch

Basic ProceduresStartingxPONENT®• On the PC desktop, click the Luminex® xPONENT® icon, or click Start > All Programs >Luminex > xPONENT >

Pagina 65 - Chapter 5: Results Page

CAUTION: Use of this software by untrained personnel can result ininaccurate data and test results. Users of xPONENT® must readthe documentation thoro

Pagina 66 - Performing Analysis

4. From the Maintenance options drop-down menu you can define the daily startup routine.Select the check box if you want to Allow running the calibrat

Pagina 67 - Current Batch Tab

NOTE: If using a strip-well plate, ensure that a strip is present in theselected well location, prior to calibration.5. Place the plate on the plate h

Pagina 68

WARNING: Correct sample probe height is critical to successful sampleacquisition and calibration. Problems with the sample probe canlead to fluid leak

Pagina 69

NOTE: See the Defining the System Initialization section for moreinformation.Calibrator xMAP beads are used to normalize the settings for the reporter

Pagina 70

3. Click Exit to exit the application.Using Online HelpEnglish-language help is available at all times while you are using xPONENT®. To displayonline

Pagina 71

Luminex® SupportViewing the Luminex® WebsiteAdditional information is available on the Luminex® website. FAQs are available at http://www.luminexcorp.

Pagina 72

Standard Terms and Conditions for Use of InstrumentProductBy opening the packaging containing this instrument product ("Product") or by usin

Pagina 73 - Saved Batches Tab

xPONENT® for FLEXMAP 3D®16

Pagina 74

General GuidelinesWARNING: Modifying or deleting xPONENT® system files can causedegradation of system performance. Repair modified or deletedxPONENT®

Pagina 75

Biological SamplesCAUTION: Human and animal samples may contain biohazardousinfectious agents. Where exposure to potentially biohazardousmaterial—incl

Pagina 76

• Maximum well volume plate = The maximum volume capacity of the wells in a selected 96and 384-well microtiter plate.NOTE: This sample volume informat

Pagina 77 - Results Subtab

xPONENT® for FLEXMAP 3D®20

Pagina 78

Samples Page OverviewSamples > SamplesUse this tab for the following:Create New Samples - Click to create a new sample.Protocol - Displays the prot

Pagina 79

Number of Samples - Displays the number of samples are in the protocol.NOTE: If a protocol is created using the same name and version as apreviously d

Pagina 80

2. In the Sample Lists section, select the protocol you are using for the sample list, thenclick Create New Samples. The Create Sample tab opens. 3. I

Pagina 81 - Log Subtab

Editing a Sample List1. Open the Samples page.xPONENT® for FLEXMAP 3D®24

Pagina 82 - Sample Details Subtab

2. In the Samples list section, choose the protocol you want to edit, then click Details. TheEdit Samples subtab opens.3. Click a sample, then use the

Pagina 83 - LIS Results Tab

hardware not provided by Luminex. If Product is purchased from a Luminex authorizedreseller, any warranty obligations shall be provided in writing dir

Pagina 84 - Reports Tab

xPONENT® for FLEXMAP 3D®26

Pagina 85

Batches Page FunctionalityBatches > BatchesOptions on the Batches tab on the Batches page are:• Create a New Batch from an existing Protocol• Creat

Pagina 86 - Generating a Report

NOTE: These tabs (except New MultiBatch) are sequential. You mustcomplete each screen in a specific order.The Pending Batches list displays the name o

Pagina 87 - Chapter 6: Protocols Page

Using the Batches Page1. Open the Batches page.2. Click one of the following:• Create a New Batch from an Existing Protocol• Create a New Batch from a

Pagina 88 - Protocol Procedures

NOTE: If the batch spans more than one plate, the tray ejects automaticallywhen all defined wells have been acquired. A dialog box opens,prompting you

Pagina 89

Use this tab to apply a kit or lot to the batch. This tab contains the following:Apply Std/Ctrl Kit - Opens the Select Std/Ctrl Kit dialog box. The d

Pagina 90

Apply Std Lot - Opens the Select Lot dialog box.Select a lot from the list and then click OK to apply the lot.Apply Values - Applies a value across or

Pagina 91

Cancel - Returns to the Batches tab.Back - Returns to the previous tab.Next - Continues to the Plate Layout tab.Plate Layout SubtabBatches > Batche

Pagina 92 - Editing a Protocol

Replicate Count - Defines a quantity of replicate sets from one to nine.NOTE: Replicate count selection must be made before adding a wellcommand.Grou

Pagina 93 - Lots and Kits Procedures

an error when attempting to run a batch on a system where theroutine does not exist.Clicking either Pre Batch Routine or Post Batch Routine opens the

Pagina 94 - Creating a Kit

Luminex's laser based fluorescent analytical test instrumentation. Buyer furtheracknowledges that the Product has not received approval from the

Pagina 95 - Protocols Tab

5. Define the settings in the Acquisition Settings section. These are Volume, XY Heater(enable/disable and set temperature), Plate Name, and enable/di

Pagina 96

16. Click Single Step to acquire the first well, then pause acquisition.17. Click Run Batch to start acquisition, or Save to save the batch for a late

Pagina 97

CAUTION: Acquiring data before the heater has reached the propertemperature can compromise test results.• Timeout - Check the Enabled box to enable th

Pagina 98

• Number of Standards - Click to type the number of standards for the protocol.• Number of Controls - Click to type the number of controls for the pro

Pagina 99

Default Analysis - The default analysis changes based on the Analysis Type selected in theSettings tab. You can change the analysis settings for all a

Pagina 100 - Plate Layout Subtab

Selected Analytes List - Selected analytes appear in a list on the right side of the analytegrid. This list includes the following information:• Name

Pagina 101

Use this tab to apply a kit or lot to the batch. This tab contains the following:• Apply Std/Ctrl Kit - Opens the Select Std/Ctrl Kit dialog box. The

Pagina 102 - Analytes Subtab

• 1:2 - Halves the standard from each previous iteration.• 1:10 (Log) - Computes a value of one-tenth of the standard from each previousiteration.• 1/

Pagina 103

Use this tab to define commands that apply to one or more wells. You can also define off-plate and maintenance commands. This tab contains the followi

Pagina 104 - Analysis Settings Dialog Box

• 123123123. . . Lays out one of each replicate set at a time in numerical order.• 111222333. . . Lays out all the replicates in a set before moving o

Pagina 105

consulting fees) that any of the Released Parties may sustain or incur as a result of any claimagainst such Released Party based upon negligence, brea

Pagina 106

• Plate Navigation - Displays a smaller plate image for the current batch.• Single Step - Instructs the system to acquire one well and then stop. Use

Pagina 107 - Chapter 7: Maintenance Page

Multi-Batch - Lists pending batches selected for the multi-batch. The list includes name and“Start at” well.Plate Layout - Opens the Multi-batch Repor

Pagina 108

2. Once you've selected a batch click Ok.3. On the next screen, select a well location for the next batch and click Add.xPONENT® for FLEXMAP 3D®4

Pagina 109 - Lot Management Tab

4. Once the batches are created, click Run.After creating a multi-batch, you can save it to the Select Pending Batch list. When saved tothis list, the

Pagina 110

4. Navigate to the location to which you want to save the file, then click Save.5. Click OK in the Export Batch dialog box. NOTE: When exporting a lar

Pagina 111 - Commands and Routines

Results PageOnce data is collected in a batch, observation and analysis take place in the Results page.This page contains the following tabs:Current B

Pagina 112

• Batch Audit• Patient Report• Protocol Reports• Protocol Settings• Plate Layout• Protocol Audit• Calibration and Verification ReportsNOTE: Calibratio

Pagina 113 - Creating a New Routine

Current Batch TabResults > Current BatchUse this tab to view results, statistics, and log information related to the current batch, and toperform s

Pagina 114 - Importing a Routine

The Current Batch tab has the following features:• Statistic - To view a particular statistic for analytes in a batch, select one of the options onthe

Pagina 115 - Probe and Heater Tab

• Analyte - Contains a list of analytes run in the batch. Select an analyte to view all statisticsfor that analyte.• Well(s) to View• Current Well - D

Pagina 116

End-User License Agreement (EULA) for Luminex®xPONENT® SoftwareThis Luminex End-User License Agreement (“EULA”) is a legal agreement between you(eithe

Pagina 117 - System Info Tab

changed to visually inspect the effect. Only the visual display of the dot plot and histogramis affected for saved batches, not the acquired data.NOTE

Pagina 118

• Date• Message• CodeLog entries indicating warnings are highlighted ion yellow. Errors are highlighted in red. Otherlog entries are not highlighted.

Pagina 119 - System Status Tab

• 3D BeadMap - The main display is the 3D BeadMap. The display shows the CL1, CL2,and CL3 axes. Left-click and drag your mouse to move the display for

Pagina 120 - Schedule Tab

• Linear• Display Position• 3D View• CL1/CL3 Plane• CL2/CL3 Plane• CL1/CL2 Plane• View• All Events• Gated Events• Show Cursors• Show Region Labels• Sh

Pagina 121 - Support Utility Tab

Use this tab to open a batch that has been run and view its details, and to export, approve, orreplay a batch.Select a Completed Batch and click Open

Pagina 122 - Sending a Support.zip File

This dialog box lets you choose the saved batches you want to display in the CompletedBatches list, based on options you select or clear in these chec

Pagina 123 - Chapter 8: Admin Page

• Plate Layout - Opens the Report dialog box, which contains the Batch Plate LayoutReport.• Approve - Opens the Batch Approval Confirmation dialog box

Pagina 124 - LIS Settings

• Recalculate data - Reanalyze the batch results using only the batch MFI values. Thebatch MFI values are recalculated using the analysis settings and

Pagina 125 - Maintenance Options

This subtab displays the following features:There are three maximize buttons in this window, one for each major pane. Click theappropriate one to maxi

Pagina 126 - Removing an Analysis Program

• Trimmed%CV of microspheres• Trimmed Standard Deviation• Trimmed Peak• %CV of Replicates• % Recovery• Expected Result - The known or expected test re

Pagina 127

SOFTWARE on additional computers, as agreed to in writing with Luminex or itsauthorized reseller, for purposes of performing ancillary tasks (i.e. pre

Pagina 128 - Group Setup Tab

• Close - Closes the batch and reopens the Saved Batches tab.• Save - Saves the batch. This button only appears if a change has been made to the batch

Pagina 129

Viewing Batch Settings1. Open the Results page, then open the Saved Batches tab.2. Click Saved Batches, then click the batch for which you want to vie

Pagina 130 - User Setup Tab

• Export - Opens the Save As dialog box to save the batch log file. Select a location andclick Save.• Close - Reopens the Saved Batches tab.Viewing Ba

Pagina 131 - Editing User Permissions

Viewing Sample Details1. Open the Results page, then open the Saved Batches tab.2. Click Saved Batches, then click the batch for which you want to vie

Pagina 132 - Create User Account Window

Others• User ID• Date• Reset• OK / Cancel• Clear - Click to turn off the filter.• Completed Samples - Displays Name, Protocol, Sample Count, Date, Sta

Pagina 133 - Edit User Account Window

• Data Interpretation• Batch Settings• Plate Layout• Batch Audit• Patient ReportIf you choose Data Interpretation, a Select Analysis section appears o

Pagina 134 - Batch Options Tab

Perfomance Verification Reports - Displays two report types. Choose either of thefollowing:• Condensed• DetailedSelect a report type, a date range fro

Pagina 135

Protocols Page FunctionalityProtocols > ProtocolsThe Protocols page enables you to create or import a new protocol, or select an existingprotocol f

Pagina 136 - Alert Options Tab

• Name• Version• Manufacturer• DateClick the Create New Protocol button or click Stds/Ctrls to move to those pages.There are action buttons at the bot

Pagina 137 - Alert Options Tasks

8. Click Next. The Analytes tab opens.9. Click the desired analyte (bead ID) in the numbered grid. Information about the analytesis displayed in a lis

Pagina 138 - CSV Options Tab

be protected by applicable copyright or other intellectual property laws and treaties. ThisEULA grants you no rights to use such content.7. EXPORT RES

Pagina 139 - Raw Run Files Export Options

NOTE: The protocol for a quantitative assay must have at least two or morestandards. The number of controls are optional.8. Click Next. The Analytes t

Pagina 140 - Archive Options Tab

• Check Mark as Intra-Well Normalization Bead if you want to use the analyte as anormalization bead.NOTE: The normalization bead is a microsphere set

Pagina 141

6. Define settings in the Analysis Settings section, selecting Qualitative as the analysistype. NOTE: The protocol for a qualitative assay contains a

Pagina 142

Exporting a Protocol1. Open the Protocols page, then open the Protocols tab.2. Select a protocol.3. Click Export. The Save as dialog box opens.4. Sele

Pagina 143 - Performing a Manual Archive

3. If the protocol uses standards, type the appropriate information for each standard in theAssay Standard Information section. In each analyte column

Pagina 144 - Licensing Tab

3. Type the name of the kit in the Name box, the lot number in the Std/Ctrl Kit Lot# box,the expiration date using MM/DD/YYYY format in the Expiration

Pagina 145

Installed Protocols - Displays installed protocols. The list includes Name, Version,Manufacturer, and Date information for each protocol.New Std/Ctrl

Pagina 146

add at least 25 µl to the sample well in addition to the sample size. The default value is 45µl.• XY heater - Select Enabled to enable the XY heater.

Pagina 147 - Report Options Tab

NOTE: Luminex recommends Fit of All Standards as the most accuratecalculation of the curve fit.Based on a range of quantitative, numerical results, a

Pagina 148

7. If you selected Lum Qual in the Formulas list, edit the values for each range underNegative, Low Positive, or High Positive. If you selected Adv Qu