91
Touring the Software
Batch Options Tab
Use this tab to set options for acquisition and batch analysis.
FIGURE 58. Batch Options Tab
This tab contains the following:
Batch Options. This section displays the following options:
• Allow running a batch if the instrument is not warmed up - Allows batches to run if
the instrument is not warmed up.
• Allow running a calibration if the instrument is not warmed up - Allows calibration
to run if the instrument is not warmed up.
• Allow running a batch with expired reagents - Allows batches to run with expired
reagents.
• Allow running a batch if XY temperature is not in range - Allows batches to run if
the XY temperature is not in range.
• Allow running a batch if the instrument is not calibrated or verified - Allows
batches to run if the instrument is not calibrated.
• Use Weighting as default option for quantitative analysis - Sets the weighting for
quantitative analysis to default to ON for new protocols.
• Warn when saving/running a New Batch with New Protocol without saving the
Protocol and/or Standard and Control information - Enables a dialog box warning
when saving or running a new batch with a new protocol when you haven’t saved the
protocol and/or standard and control information.
• Prompt for reacquiring of the last errored wells for an aborted partial batch -
Allows users to choose to reacquire aborted wells when resuming a stopped/partial
batch or to start with the next unacquired well.
Batch Settings. This section displays the following options: