Luminex xPONENT 4.0 Software Manual
60
FIGURE 40. Analysis Settings Dialog Box
If you selected Quantitative on the Settings tab, the default analysis formula is 5P
weighted. To change the default, select one of the following from the Method list:
No Analysis
Cubic Spline
Linear Fit
Logistic 4P
Logistic 5P
If you select Luminex Logistic 4P or Luminex Logistic 5P, select a weight type of either
None or 1/Y
2
.
If you selected Qualitative on the Settings tab, the default analysis is Luminex
Qualitative. Change the default value by selecting either Lum Qual or No Analysis.
Click Apply to All Analytes to apply your selection to all selected analytes. Click OK to
change the default analysis to the analysis you selected. Click Cancel to close the dialog
box without saving. For more information, see “Analysis Settings” on page 25.
Units. Type the desired units for the analytes in this box.
Count. Type the desired bead count for the analytes by clicking in the Count box. If each
selected bead set does not acquire this number of events, a warning is added to the log
that not enough bead events were acquired. If you select bead sets that are not present,
the analyzer continues to acquire, trying to reach the number of events per bead for bead
sets that are not in the sample. Therefore, choose only the bead sets present in your
sample.
Apply All. Applies the information in the Units and Count boxes to all analytes.