6. Click Run. The Routine Message dialog box opens when the routine is complete.
7. Click OK.
Importing a Routine
1. Open the Maintenance page, then open the Cmds & Routines tab.
2. Click Import.
3. In the Open dialog box, browse for the file you want to import, then click Open. The
routine becomes the active routine.
Exporting a Routine
1. Open the Maintenance page, then open the Cmds & Routines tab.
2. Click Export.
3. In the Save As dialog box, navigate to the folder where you want to store the routine file,
then click Save.
NOTE: The export option is only available for routines that have been saved.
Probe and Heater Tab
Maintenance > Probe & Heater
Use this tab to adjust the probe height and plate heater settings.
Maintenance Page
113