Luminex FLEXMAP 3D Hardware User Manual Manual de usuario Pagina 57

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 72
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 56
Symptom Description Possible Problem Solution
xMAP
microspheres do
not form a
cohesive group
and are
somewhat
scattered.
There is air in the
instrument.
Verify the sample probe height. Use the
software to run a prime command three
times, an alcohol flush command twice,
then a wash command three times with
distilled water.
xMAP
microspheres are
widely scattered.
The sheath fluid is
empty.
Make sure there is sheath fluid in the
sheath container. Use the software to
run a prime command until all air is
pushed out of the instrument.
FLEXMAP 3D
®
Hardware User Manual
46
Vista de pagina 56
1 2 ... 52 53 54 55 56 57 58 59 60 61 62 ... 71 72

Comentarios a estos manuales

Sin comentarios