PN 89-00002-00-150 Rev. A C - 1
Luminex Sheath Delivery System
Overview Use this appendix to install, use, and maintain the Luminex Sheath
Delivery System (Luminex SD™).
Specifications
General Ambient temperature: 15°C to 30°C (59° to 86°F)
Humidity: 20% to 80%, noncondensing
Altitude: designed to operate at up to 2400m (7874 feet) above
mean sea level
Physical dimensions: 20 cm (8 inches) W x 30 cm (11.75 inches)
D x 24.75 cm (9.75 inches) H
Weight: 9 kg (20 lbs)
UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1 z
Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1
Power Input voltage range: 100-240 V~, 0.4 Amps, 47-63 Hz
AC inlet fuse: 2 Amp, 250 V~, Time lag
Intended Use With the Luminex SD system, you can run samples continuously
without refilling sheath bottles. The Luminex SD automatically
draws sheath from a non pressurized bulk sheath container to
constantly maintain a reservoir of pressurized sheath fluid. A single
20 liter sheath container provides enough fluid for 48 hours or more
of normal operation.
USE OF NON-LUMINEX APPROVED SHEATH FLUID
SHALL CONSTITUTE “IMPROPER USE” AND MAY VOID