Luminex 200 User Manual for LDS Version 1.7 xMAP Technology
3 - 4 PN 89-00002-00-150 Rev. A
• AC inlet fuse: 3 Amp, 250 V~, fast acting
Optics • Reporter laser: 532 nm, nominal output 10-15 mW, maximum
500 mW, frequency-doubled diode; mode of operation,
continuous wave (CW)
• Classification laser: 635 nm, 9.1 mW ± 6%, maximum output 25
mW, diode; mode of operation, continuous wave (CW)
• Reporter detector: Photomultiplier tube, detection bandwidth of
565-585 nm
• Classification detector: Avalanche photo diodes with temperature
compensation
• Doublet discrimination detector: Avalanche photo diodes with
temperature compensation
Fluidics • Sheath flow rate 90 µL ± 5 µL/second
• Cuvette: 200 micron square flow channel
• Sample injection rate: 1 µL/second ± .05 µL/second
• Sample uptake volume: 20-200 µL
Electronics • Reporter channel detection: A/D resolution 14 bits
• Communications interface: USB
Luminex XYP
Specifications
• Ambient temperature: 15°C to 30°C (59°F to 86°F)
• Humidity: 20% to 80%, noncondensing
• Altitude: designed to operate at 2400 m (7874 ft.) above mean
sea level or below.
• Physical dimensions: 44 cm (17.25 in) W x 61 cm (24 in) D x 8
cm (3 in) H
• Weight: 15 kg (33 lbs)
• UL installation category: UL Installation Category II, as defined
in Annex J of UL 3101-1.
• Pollution degree: UL Pollution Degree 2, as defined in section
3.7.3.2 of UL 6101A-1
• Heater operating range: 35°C to 60°C (95°F to 140°F) with
tolerance 0°C to +2°C
• Input voltage range: 100-240 V~ ± 10%, 0.4 Amps, 47-63 Hz
• AC inlet fuse: 2 Amp, 250 V~, time lag
• Communications interface: RS 232
• Shipping and storage: allowable shipping and storage
temperature and humidity ranges, 0°C to +50°C and 20%-80%
noncondensing, respectively